Current Issues of Pharmacy and Medical Sciences

Desloratadine analysis: as a pharmaceutical preparation and after accelerating ageing

Curr Issues Pharm Med Sci., Vol. 28, No. 3, Pages 181-185

Katarzyna Bober, Marlena Płonka, Marek Miszczyk

1 Department of Analytical Chemistry, Faculty of Pharmacy, Medical University of Silesia in Katowice, Jagiellonska 4, 41-200 Sosnowiec, Poland
2 Institute of Plant Protection – National Research Institute, Pesticide Quality Testing Laboratory, Gliwicka 29, 44-153 Sosnicowice, Poland


DOI: 10.1515/cipms-2015-0068

Abstract

Desloratadine is a biologically active compound that is not described in the Polish Pharmacopoeia IX, hence, its study is particular important. The aim of this work was to establish a procedure for desloratadine analysis by way of HPTLC in combination with densitometry, so as to be able to ascertain its presence and degree of presence within pharmaceutical preparations. In our work, a mixture of ethyl acetate, n-butanol, ammonia and methanol was used as the mobile phase. Moreover, HPTLC plates pre-coated with silica gel 60F254 were also employed. The proposed method was tested and subsequently validated. Spectrodensitometric analysis was then performed to determine the optimal wavelength for the quantitative determination (λ=276 nm), and following this, a quantitative analysis of desloratadine within certain pharmaceutical preparations was performed. Our research also took into consideration an analysis of the products of desloratadine decomposition that come about as a result of the accelerated aging of its solutions. The employed procedure for accelerating the aging of such desloratadine solutions consisted of heating these at 40°C and then irradiating the solution surfaces with UV light. The changing color of these solutions after 2 hours of exposure served to indicate that degradation had occurred. Of note: as a result of irradiation with UV light, desloratadine content was seen to decrease with time, declining to almost zero after 30 hours. However, heating a solution of desloratadine alone did not induce a change in its content. Solutions of desloratadine that had previously undergone irradiation and heating were also analyzed to ascertain whether new substances were present. For this purpose, the GC-MS process was employed. As a result of this procedure, the spectrum of the solution after aging showed the presence of several new peaks that displayed retention several times larger and smaller than the normal desloratadine peak.

Full text

Keywords

desloratadine, densitometry, HPTLC, GC-MS, accelerating ageing.

Calendar

December 2019

Mon Tue Wed Thu Fri Sat Sun
            01
02 03 04 05 06 07 08
09 10 11 12 13 14 15
16 17 18 19 20 21 22
23 24 25 26 27 28 29
30 31